Abstract

The application of NMR Fricke-gelatin dosimetry to the imaging of dose distributions around high dose rate (HDR) radioactive Ir-192 seeds in brachytherapy is presented. The duration of the irradiations required to give changes of the Fricke-gelatin NMR characteristics, which were readily detected by the magnetic resonance imager, typically ranged from 4 to 25 min for single source and multiple source (33 source positions distributed among three catheters) geometries, respectively. A fast approach for determining radiation dose from MR image intensity using a calibration curve is described. NMR Fricke-gel dosimetry is shown to be a viable means of verifying the dose distributions from brachytherapy sources.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.