Abstract
Assessing peripheral vasculature health has the potential to impact clinical decision making in terms of treating patients with cardiovascular disease. The electrical conductivity of certain tissue regions within the forearm change as blood vessels undergo pulsatile dilation in synchrony with the beating of the heart. We use dynamic electrical impedance tomography (EIT) gated to the peak of a pulse oxymetry plethysmography waveform to image this temporally varying spatial conductivity. A phantom imaging experiment is presented showing that small conductivity changes of less than 1 mm are detectable using the developed dynamic EIT system. This system is used to image a volunteer's forearm during resting cardiovascular activity. Similar structures are observed in the plethysmography trace and the temporally varying conductivity. Spectral analysis shows that the maximum amplitude is occurring at frequencies of 1.19 Hz and 1.21 Hz for the plethysmography trace and conductivity trace, respectively. This preliminary data suggests that EIT may be sensitive enough to visualize cardiac-based pulsatility in the peripheral vessels of the forearm.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.