Abstract

The emission of Cherenkov photons from human and animal tissue can be observed during clinical x-ray or particle beam irradiation. However, imaging this weak emission with the necessary single-photon sensitivity in the clinical room is challenging because of milliwatt-level ambient room lighting and the presence of stray high-energy radiation. In this Letter, we demonstrate, to the best of our knowledge, the first Cherenkov imaging with a time-gated quanta image sensor employing a large single-photon avalanche diode (SPAD) array. Detecting single Cherenkov photons was possible with high photon avalanche gain, fast temporal gating, and moderately high ∼7% photon detection probability. Single-bit digitization and active SPAD quenching enabled stray x-ray noise suppression and photon-noise-limited imaging in a clinical environment. This type of imaging allows the knowledge of location, shape, and surface dose of the therapeutic beam radiotherapy with the stability of solid state-based detection.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.