Abstract

We report the use of ultrasound modulated optical tomography (UOT) with heterodyne parallel detection to locally sense and image blood flow deep inside a highly scattering medium. We demonstrate that the UOT signal is sensitive to the speed of the blood flow in the ultrasound focus and present an analytical model that relates UOT signals to the optical properties (i. e. scattering coefficient, anisotropy, absorption, and flow speed) of the blood and the background medium. We found an excellent agreement between the experimental data and the analytical model. By varying the integration time of the camera in our setup, we were able to spatially resolve blood flow in a scattering medium with a lateral resolution of 1.5 mm.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call