Abstract

A frequently reported drawback of ethylene vinyl alcohol copolymer-based liquid embolic agents is the production of artifacts in diagnostic imaging. New embolic agents, such as Precipitating hydrophobic injectable liquid (PHIL; MicroVention, Tustin, CA, USA), are supposed to induce significantly fewer artifacts. The purpose of this study is to assess the degree of artifacts induced by the liquid embolic agents Onyx (Medtronic Neurovascular, Irvine, CA, USA) and PHIL in conventional computed tomography (CT), cone-beam CT and magnetic resonance imaging (MRI) in an experimental invivo model. In 10 pigs the rete mirabile was embolized with Onyx ( n = 5) or PHIL ( n = 5). After embolization, conventional CT, cone-beam CT and MRI were performed. The degree of artifacts was graded qualitatively (five-point scale; for CT and MRI) and quantitatively (HUs of well-defined regions of interest (ROIs); for CT only). Artifacts were significantly more severe for Onyx both in the qualitative (e.g. conventional CT: 2 versus 5 (medians); p = 0.008) and in the quantitative image analysis (e.g. cone-beam CT: standard deviation of a ROI near to the embolic agent cast, 94 HU versus 38 HU (medians); p = 0.008). Neither Onyx nor PHIL produced any apparent artifacts in MRI. PHIL produces fewer artifacts than Onyx in conventional CT and cone-beam CT in an experimental invivo model.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.