Abstract

Imaging of the focal spot size in X-ray generators can be achieved by means of a pinhole in a highly attenuating material, such as gold. For microfocus generators with spot sizes of only around 10 microns or less, the material must be very thin to avoid an impractical aspect ratio. With a 90 kV source, only 11% attenuation is attained with 5 µm gold. For a pinhole that is smaller than the focal spot, the signal-to-noise ratio (SNR) may be less than 1. To image the focal spot of a medical X-ray generator, a coded aperture has been used previously to reduce the exposure time required, however the same technique does not appear to have been used to increase the SNR when the attenuation is very low. Such a method is used here, using a no-two-holes-touching variation of a modified uniformly redundant array (MURA). In a prototype sample, with only 5 µm gold having 2.75 µm holes, the focal spot of a microfocus X-ray generator used in a micro-CT system could be clearly visualised and quantified. Directionality of the aberrations made focussing of the X-ray spot more intuitive and reduced the time required to obtain an optimal, quantifiable focus.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.