Abstract

We sought to determine the accuracy of an electromagnetic image guidance surgical navigation system in localizing the midpoint of the internal auditory canal (IAC) and other structures of the temporal bone through the middle cranial fossa approach. Seven fresh cadaveric whole heads were dissected via a middle cranial fossa approach. High-resolution CT scans were used with an InstaTrak 3500 Plus electromagnetic image guidance system (General Electric, Fairfield, CT). We evaluated the accuracy of identifying several middle cranial fossa landmarks including the midpoint of the IAC; the labyrinthine segment of the facial nerve; and the arcuate eminence, the carotid artery, and foramen spinosum. We were able to identify the middle of the IAC within 2.31 mm (range 0.65-7.52 mm, SD 2.39 mm). The arcuate eminence could be identified within 1.86 mm (range 1.49-2.37 mm, SD 0.36 mm). We noted some interference when the handpiece was within 6 to 8 cm of the microscope. Although computer-aided navigational tools are no substitute for thorough knowledge of temporal bone anatomy, we found the InstaTrak system reliable in identifying the midpoint of the IAC to within 2.4 mm through a middle fossa approach.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.