Abstract
In Nondestructive Evaluation (NDE), ultrasound imaging is a useful tool for detecting flaws or measuring corrosion in metal pipes. Phased-array systems are becoming standard in NDE, as well as post-processing imaging techniques such as the Total Focusing Method. These techniques require knowledge of parameters such as distance between transducer elements, times of arrival and sound speed to produce sharp images. A deviation in sound speed, for instance, may compromise image accuracy. In this paper, we propose a sound speed estimator based on image quality metrics. The approach consists in generating several reconstructed images from the same acquisition signals, but using a set of candidate sound speeds. The speed value that yields the sharpest image is then chosen as the estimated sound speed. Estimates made on data obtained by simulation of a typical scenario presented relative combined uncertainties as low as 0.0166%. The results, obtained in both simulated and real scenarios, show good agreement with expected speeds. The proposed method can be considered a reference-free sound speed calibration method aimed at field applications.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.