Abstract
Image restoration to deblur smoothing caused by the finite-size X-ray beam profile for a simulated computed tomography (CT) system is presented. Three simple image restoration methods are compared when the point-spread-function (PSF) is spatially invariant. In the first restoration method, an iterative least squares solution, regularized with the image norm and constrained by the boundary of the object, is obtained from the projection data. In the second method, a Wiener filter, designed using the power spectrum of CT noise, is applied to the reconstructed CT image. The third method obtains a weighted least-squares solution, by iteration, from the reconstructed CT image; the solution is regularized with the weighted image norm. Restored images were compared with the image obtained using filtered backprojection method. Differences between these images were evaluated qualitatively.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.