Abstract
To evaluate the image quality of virtual monochromatic images (VMI) reconstructed from dual-energy dual-source noncontrast head CT with different reconstruction kernels. Twenty-five consecutive adult patients underwent noncontrast dual-energy CT. VMI were retrospectively reconstructed at 5-keV increments from 40 to 140 keV using quantitative and head kernels. CT-number, noise levels (SD), signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) in the gray and white matter and artifacts using the posterior fossa artifact index (PFAI) were evaluated. CT-number increased with decreasing VMI energy levels, and SD was lowest at 85 keV. SNR was maximized at 80 keV and 85 keV for the head and quantitative kernels, respectively. CNR was maximum at 40 keV; PFAI was lowest at 90 (head kernel) and 100 (quantitative kernel) keV. Optimal VMI image quality was significantly better than conventional CT. Optimal image quality of VMI energies can improve brain parenchymal image quality compared to conventional CT but are reconstruction kernel dependent and depend on indication for performing noncontrast CT.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.