Abstract

We evaluated the potential of prospectively ECG-triggered high-pitch spiral acquisition with low tube voltage and current in combination with iterative reconstruction to achieve coronary CT angiography with sufficient image quality at an effective dose below 0.1 mSv. Contrast-enhanced coronary dual source CT angiography (2 × 128 × 0.6 mm, 80 kV, 50 mAs) in prospectively ECG-triggered high-pitch spiral acquisition mode was performed in 21 consecutive individuals (body weight <100 kg, heart rate ≤60/min). Images were reconstructed with raw data-based filtered back projection (FBP) and iterative reconstruction (IR). Image quality was assessed on a 4-point scale (1 = no artefacts, 4 = unevaluable). Mean effective dose was 0.06 ± 0.01 mSv. Image noise was significantly reduced in IR (128.9 ± 46.6 vs. 158.2 ± 44.7 HU). The mean image quality score was lower for IR (1.9 ± 1.1 vs. 2.2 ± 1.0, P < 0.0001). Of 292 coronary segments, 55 in FBP and 40 in IR (P = 0.12) were graded "unevaluable". In patients with a body weight ≤75 kg, both in FBP and in IR, the rates of fully evaluable segments were significantly higher in comparison to patients >75 kg. Coronary CT angiography with an estimated effective dose <0.1 mSv may provide sufficient image quality in selected patients through the combination of high-pitch spiral acquisition and raw data-based iterative reconstruction.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call