Abstract

The purpose of this study was to investigate the effects of an organ-based tube-current modulation (OBTCM) system on image quality and age-specific dose in head and chest CT examinations. Image noise, contrast-to-noise ratio (CNR) and image entropy were assessed using statistical and entropy analyses. Radiation doses for newborn, 6-y-old child and adult phantoms were measured with in-phantom dosimetry systems. The quality of CT images obtained with OBTCM was not different from that obtained without OBTCM. In head CT scans, the eye lens dose decreased by 20-33 % using OBTCM. In chest CT scans, breast dose decreased by 5-32 % using OBTCM. Posterior skin dose, however, increased by 11-20 % using OBTCM in head and chest CT scans. The reduction of effective dose using OBTCM was negligibly small. Detailed image quality and dose information provided in this study can be effectively used for OBTCM application.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.