Abstract

A robotic image-guided radiosurgical system has been modified to treat extra-cranial sites using implanted fiducials and skeletal landmarks to locate the treatment targets. The system has been used to treat an artero-venous malformation in the cervical spine, a recurrent schwannoma of the thoracic spine, a metastatic adenocarcinoma of the lumbar spine, and three pancreatic cancers. During each treatment, the image guidance system monitored the position of the target site and relayed the target coordinates to the beam-pointing system at discrete intervals. The pointing system then dynamically aligned the therapy beam with the lesion, automatically compensating for shifts in target position. Breathing-related motion of the pancreas lesions was managed by coordinating beam gating with breath-holding by the patient. The system maintained alignment with the spine lesions to within ± 0.2 mm on average, and to within ± 1 mm for the pancreatic tumors. This experience has demonstrated the feasibility of using image-guided robotic radiosurgery outside the cranium.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.