Abstract

We validated a navigator-echo-triggered sequence that drives magnetization before cardiac-gated inversion recovery T1 turbo field echo acquisition, in the sedated free-breathing pediatric population. Cardiac magnetic resonance imaging was performed on sedated infants with single ventricle. We calculated the signal-to-noise ratios and contrast-to-noise ratios of 2 groups of images obtained using respiratory triggering with and without navigator echo. All images were then visually assessed by 2 observers. The signal-to-noise ratio and the contrast-to-noise ratio were significantly higher with than without navigator echo (p<0.01; p<0.05). The visual assessment scores were also consistently better with than without navigator echo (p<0.01). Free-breathing navigator echo was found to have the advantage of decreasing the motion artifact caused by respiration. Cardiacgated inversion recovery T1 turbo field echo sequence for free-breathing navigator-echo-triggered respiration allows for the acquisition, in sedated infants, of diagnostic images whose quality exceeds that of the non-navigator-echo-triggered alternative.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.