Abstract

We have studied effects of MR (magnetic resonance) image distortion on polymer gel dosimetry of Gamma Knife stereotactic radiosurgery systems. MR images of BANG polymer gel phantoms were acquired by using a Hahn spin-echo sequence and a fast 3D imaging GRASS sequence. Image artifacts were studied by varying the directions of frequency encoding and the receiver bandwidth. The phantoms were also CT (computed tomography) scanned. The studies showed that the measured dose distributions are shifted by 1.8+/-0.5 mm (2 pixels) in the frequency encoding direction. The magnitude of the shift is inversely proportional to the receiver bandwidth in agreement with theory. Comparison of MRI with CT showed the same image shift. We concluded that the discrepancy is caused by MR image distortion due to a difference in susceptibility effects between the phantom and the fiducial markers of the Leksell localization box.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.