Abstract
There is no consensus yet on the best protocol for planar image acquisition and interpretation of radiolabelled white blood cell (WBC) scintigraphy. This may account for differences in reported diagnostic accuracy amongst different centres. This was a multicentre retrospective study analysing 235 WBC scans divided into two groups. The first group of scans (105 patients) were acquired with a fixed-time acquisition protocol and the second group (130 patients) were acquired with a decay time-corrected acquisition protocol. Planar images were interpreted both qualitatively and semiquantitatively. Three blinded readers analysed the images. The most accurate imaging acquisition protocol comprised image acquisition at 3 - 4 h and at 20 - 24 h in time mode with acquisition times corrected for isotope decay. Using this protocol, visual analysis had high sensitivity and specificity in the diagnosis of infection. Semiquantitative analysis could be used in doubtful cases, with no cut-off for the percentage increase in radiolabelled WBC over time, as a criterion to define a positive scan.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: European Journal of Nuclear Medicine and Molecular Imaging
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.