Abstract

A calibration facility and associated measurement procedures have been established at the National Institute of Metrology (NIM, CHINA) to provide high illuminance calibrations in the range from 1 klx to 200 klx. An adjustable high illuminance source based on flux addition of LEDs in an integrating sphere has been developed, and the characteristics of the high illuminance source have been tested and analyzed. A correction method for the issues caused by the inherent structure of the light source has been proposed. A photometer was calibrated using the superposition method in the range (1–200) klx with a relative uncertainty from 0.5% to 1.0% varying with range (coverage factor k = 2).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.