Abstract

Absolute values of work functions can be determined in ultraviolet photoemission spectroscopy (UPS) by measuring the minimum kinetic energy of secondary electrons generated by a known photon energy. However, some samples can produce spectra that are difficult to interpret due to additional intensity below the true secondary electron cutoff. Disordered absorbates on elemental metals add small intensity below the onset for the transition metal surfaces studied, which can be attributed to energy losses after photoelectrons are generated. In contrast, spectra from WO3−x films can produce multiple onsets with comparable intensity which do not fit this model. False onsets (in the context of work function measurements) can be minimized by optimizing experimental detection parameters including limiting analyzer acceptance angles and pass energy. True work functions can be identified by examining the onsets as the sample bias is varied.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.