Abstract
Two interference peaks which generally appeared in company with 13C labeled 2,3,7,8-tetrachlorodibenzofuran (13C12-2,3,7,8-TCDF) in the same ion channel during dioxin analysis for biological samples were identified using high resolution gas chromatography/high resolution mass spectrometry (HRGC/HRMS) and high resolution gas chromatography/low resolution mass spectrometry (HRGC/LRMS). It was firstly inferred that the interference peaks should be the two isomers of dichlorodiphenyldichloroethylene (DDE), which was a breakdown product of dichlorodiphenyltrichloroethane (DDT), one of the typical organic chlorine pesticides (OCPs). Thereafter, the standard solution of DDE including o,p'-DDE and p,p'-DDE was analyzed for confirmation. By evaluation of the peak separation in HRGC/HRMS, comparison of the GC retention times and ion abundance ratios of the two interference peaks in real samples with the two DDE isomers in standard solution, the interference peaks were finally confirmed as o,p'-DDE and p,p'-DDE in sequence on a DB-5MS column. This study provided valuable information for accurate identification of dioxin compounds during the biological sample analysis.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.