Abstract

The absence of tactile force information at the tip of a medical instrument makes precise minimally invasive surgery difficult and is prone to resulting in serious outcomes. In this paper, a fiber optic force sensor is fabricated based on Fabry-Perot interferometry and integrated with a puncture needle. The force sensor is attached to the needle tip, where interactive force arises as it inserts into soft tissue. Needle insertion experiments have been conducted on ex vivo swine liver and belly with the calibrated force sensor. Using wavelet transform method, the acquired force data are analyzed and used to identify layered tissue types and boundaries. We found that the force amplitudes are not always identical, but the patterns of forces are almost the same, which enables us to identify layered tissues and to realize a safe needle insertion procedure under robot control.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.