Abstract

Stable complexes are required during the ion chromatographic (IC) separation of Fe-polycarboxylic acid complexes. Electrospray ionization mass spectrometry (ESI-MS) was used to identify 1:1 stoichiometric complexes of Fe[HEDTA], Fe[EDTA](1-) and Fe[DTPA](2-), and the spectra showed that these Fe complexes were stable in solution. Furthermore, inductively coupled plasma mass spectrometry (ICP-MS) using an octopole reaction system (ORS) reduced polyatomic ion (40)Ar(16)O(+) interference in the detection of (56)Fe via the addition of either H(2) or He to the ORS, with He at a flow rate 3.5 mL min(-1) being the optimum collision gas. Finally, IC/ICP-MS was used for the separation and detection of Fe complexes with an eluent containing 30 mM (NH(4))(2)HPO(4) at pH 8.0, but only Fe[HEDTA], Fe[EDTA](1-) and Fe[DTPA](2-) were observed within 10 min with reasonable resolution. Detection limits in the range of 10-13 µg L(-1) were achieved using He as the collision gas. The proposed method was used for the determination of Fe species in soil solutions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.