Abstract

In high intensity focused ultrasound (HIFU) treatment, it is crucial to accurately identify denatured and normal biological tissues. In this paper, a novel method based on compressed sensing (CS) and refined composite multi-scale fuzzy entropy (RCMFE) is proposed. First, CS is used to denoise the HIFU echo signals. Then the multi-scale fuzzy entropy (MFE) and RCMFE of the denoised HIFU echo signals are calculated. This study analyzed 90 cases of HIFU echo signals, including 45 cases in normal status and 45 cases in denatured status, and the results show that although both MFE and RCMFE can be used to identify denatured tissues, the intra-class distance of RCMFE on each scale factor is smaller than MFE, and the inter-class distance is larger than MFE. Compared with MFE, RCMFE can calculate the complexity of the signal more accurately and improve the stability, compactness, and separability. When RCMFE is selected as the characteristic parameter, the RCMFE difference between denatured and normal biological tissues is more evident than that of MFE, which helps doctors evaluate the treatment effect more accurately. When the scale factor is selected as 16, the best distinguishing effect can be obtained.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call