Abstract

A GC–UV instrumental set up with two different GC units has been used for determination of specific functional groups and compounds in complex mixtures. Separations have been made using a micro gas chromatograph built into a gas flow cell and by means of an external capillary gas chromatograph linked to the same gas flow cell. Four various applications (cigarette smoke, petroleum, dust, flavour) have been performed in order to demonstrate the potential of the GC–UV method. Gas phase UV spectra have been recorded in the region of 168–330 nm. Based on a gas phase spectrum reference library the identification of unknowns as well as the determination of specific functional groups have been achieved. A table showing the spectral shapes and positions of the absorption bands for 50 specific functional groups is presented. The advantage of using derivative spectra in order to amplify spectral details and improve selectivity is discussed. Regarding sensitivity, it has been found that identifications can be made in the mid-pg range and limit of detection for naphthalenes are at a level of 0.5–3 pg/s.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.