Abstract

PurposeThe performance of the AgilityTM multileaf collimator was investigated with a focus on dynamic, small fields for flattening filter free (FFF) beams.MethodsIn this study we have developed a simple tool to test the robustness of the control mechanisms during dynamic beam delivery for Elekta’s VersaHD linear accelerator with Integrity 4.0.4 control software. We have programed the planning system to calculate dose for delivery of sweeping gaps. These sweeping gaps have a constant speed, constant size, and are delivered at a constant dose rate. Therefore they specifically identify delivery problems in dynamic mode.ResultsThe Elekta AgilityTM control mechanism fails to maintain accurate delivery for small, dynamic sweeping gaps. For small gap sizes, the AgilityTM control mechanism delivers a field that is more than four times the size of the planned field width without generating an interlock. This has dosimetric implications: The discrepancy between calculated and measured doses increases with decreasing gap size and exceeds 10% and 60% at isocenter for a 3.5 mm and 1 mm gap size, respectively.ConclusionA deficiency of the AgilityTM control system was identified in this study. This deficiency is a potential source of error for volumetric modulated arc therapy fields and could therefore contribute to relatively high failure rates in quality assurance measurements, especially for FFF beams.