Abstract

Genotyping errors can perturb estimates of genetic diversity and adversely affects the reliability of forensic identification, which has become increasingly recognized. During our daily laboratory work, we hit upon a discordant short tandem repeat (STR) genotype at the D2S1338 locus between Identifiler Plus®16 system kit and Power Plex®21 system kit. To figure out the reason of the discordance, nested PCR was performed to amplify the D2S1338 locus and then the final amplicons were sequenced. We found a G to G/A transition at the position on which caused an allele dropout of Identifiler Plus®16 kit.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.