Abstract
A wide variety of organic solvents within flammable hazardous wastes were determined by GC/FT-IR using external calibration. The hazardous waste samples were prepared for analysis by dilution with hexane. The type of compounds analyzed included acetate esters, alcohols, aromatics, ethylene glycol ethers, chlorinated hydrocarbons, ketones, and other industrial solvents. The instrumental detection limits for the various compounds ranged from 4 to 240 μg/mL. Calibration curves were linear over one to two orders of magnitude and stable for several months. The percent relative standard deviation ranged from 0.9 to 10.0% and the percent bias was within ±23% for all analytes.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.