Abstract
Appropriate ICD programming is the key to prevent inappropriate shock delivery, that is closely associated to a negative patients' outcome. Review of the literature on ICD therapy to generate ICD programmings that can be applied to the broad population of ICD and CRT-D carriers. Arrhythmia detection should occur with a detection time ranging 9″-12″ in the VF zone, and 15″-60″ in the VT zone. Discriminator should be applied at least up to 200 bpm. ATP therapy is applied to all VTs up to 250 bpm, with a success rate of 70%. Inappropriate shocks should occur in <3.6% of patients. Tailored ICD programming can be achieved following evidence from large ICD trials. Pre-defined settings that are saved on the programmer and that can be uploaded at device implantation help to ensure optimal programming and to avoid random errors.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.