Abstract

The IBA myQA® SRS high-resolution solid-state detector was evaluated in the context of robotic radiosurgery delivered using CyberKnife®. The performance was investigated for periodic machine delivery quality assurance (DQA) and patient-specific treatment verification. myQA® SRS is a 140 × 120 mm CMOS matrix with 400 µm resolution, allocated in a cylindrical ABS phantom topped by a hemispheric cap. Evaluations included: periodic DQA tests, angular response, dose-rate dependence and Iris variable aperture collimator field size measurements. For patient-specific QA various intracranial targets were studied (Gamma Index, 3%/1 mm agreement criteria), taking into account also the detector’s angular response. Results for periodic DQA were in accordance with the machine commissioning data. Dose-rate dependence was confirmed, and angular response tests resulted in a signal decay >5% when beams were delivered outside a ±50° amplitude cone with respect to the vertical direction. Concerning patient-specific QA, >50° angled beams elimination led to a Gamma Index passing rates improvement ranging between +3% and +115%. IBA myQA® SRS proved to be a suitable device for many CyberKnife® constancy DQA checks, providing high-resolution real-time results. Patient-specific Gamma tests showed high passing rates once angular dependence corrections were performed, even in high complexity treatments such as those for trigeminal neuralgia targets.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call