Abstract

ObjectiveTo describe a series of cases of iatrogenic retinal breaks (IRBs) caused by the infusion fluid flow of a 25-gauge pars plana vitrectomy (PPV) system. DesignRetrospective case series. MethodsDuring 25-gauge PPV, 4 cases had IRBs caused by infusion fluid flow. The IRBs rapidly progressed to localized retinal detachment. ResultsThe first 3 cases had IRBs on the nasal quadrant midperiphery of the retina. The IRBs were treated with laser retinopexy and tamponade during surgery. Case 4 had a macular hole and macular detachment during scleral indentation. The IRBs seemed to be caused by intraocular pressure (IOP) control mechanisms of the vitrectomy device. ConclusionsTo prevent IRBs caused by infusion fluid flow, we recommend using an IOP control limit of 4 mL/min for 25-gauge vitrectomy, with valved cannulas. In addition, the surgeon must be cautious during scleral indentation and air-fluid exchange not to cause a rebound hypotonia.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.