Abstract

To report a case of a patient who suffered a full thickness macular hole (FTMH) due to the accidental utilization of the selective laser trabeculoplasty (SLT) mode of a dual mode laser. Case report. A 69-year-old woman experienced vision loss in her left eye immediately after undergoing attempted Neodymium:yttrium-aluminum-garnet (Nd:YAG) capsulotomy using a Nd:YAG-SLT laser system. Post-injury visual acuity was 20/25 and 20/800 in the right and left eyes, respectively. Ophthalmic exam and multimodal imaging revealed multiple macular hemorrhages and an irregular FTMH. The patient required multiple surgeries including an autologous retinal transplant to achieve hole closure. Macular hole formation is a devastating consequence of inadvertent use of the SLT mode when performing a Nd:YAG laser capsulotomy with a Nd:YAG-SLT laser system. We emphasize the importance of ensuring the correct mode is selected for treatment to minimize the risk of retinal damage and permanent vision loss.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.