Abstract
Bioprosthetic valve dysfunction (BVD) is typically a progressive process related to natural wear of the prosthesis. With acute presentations, possible durability issues or iatrogenic causes need to be considered. Here, we present 2 patients with acute BVD of self-expanding, transcatheter aortic valve replacement post-heart catheterization. The presentations and outcomes, in otherwise normally functioning valves antecedent to the heart catheterizations, raise the question of the increased complexity of coronary access in this valve platform, and whether that or other features provide for greater risk of such events. We believe this to be the first publication of such events and they help to highlight the importance of valve implantation planning, as well as familiarity with the potential complexity of coronary access during heart catheterization.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.