Abstract

Independent dosimetry audits are an important intervention in radiotherapy for quality assurance. Electron beams, used for superficial radiotherapy treatments, must also be tested in dosimetry audits as part of a good quality assurance program to help prevent clinical errors. To establish a new service for IAEA/WHO postal dosimetry audits in electron beams using RPL dosimeters. A novel postal audit methodology employing a PMMA holder system for RPLDs was developed. The associated correction factors including holder dependence, energy dependence, dose response non-linearity, and fading were obtained and tested in a multi-center (n=12) pilot study. A measurement uncertainty budget was estimated and employed in analyzing the irradiated dosimeters. Holder and energy correction factors ranged from 1.004 to 1.010 and 1.019 to 1.059 respectively across the energy range. The non-linearity and fading correction models used for photon beams were tested in electron beams and did not significantly increase measurement uncertainty. The mean dose ratio±SD of the multi-center study was 1.001±0.011. The overall uncertainty budget was estimated as±1.42% (k=1). A methodology for IAEA/WHO postal dosimetry audits in electron beams was developed and validated in a multi-center study and is now made available to radiotherapy centers as a routine service.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.