Abstract

In the modern anesthesia machine, there are various safety checks in place to help prevent the delivery of a hypoxic gas mixture to the patient. They include the pin index safety system (PISS), diameter index safety system (DISS), failsafe valve, oxygen-nitrous oxide proportioning system, oxygen supply failure alarm, flowmeter sequence, and, most distally, the oxygen analyzer. The PISS is a feature in the high-pressure system. The DISS, failsafe valve, and oxygen failure alarm are in the intermediate-pressure system. The flowmeters, proportioning system, and oxygen analyzer are in the low-pressure system. This chapter undertakes a discussion of the distinct role of each feature and their limitations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.