Abstract
To assess the effect of the perfusion pressure (PP) during machine perfusion (MP) on the preservation quality of kidney grafts, we compared mean PPs of 25 and 30mmHg using a porcine autotransplantation model. After assessment of the microcirculation, animals underwent left nephrectomy. Thereafter, kidneys were washed out followed by 20h of MP at 25mmHg (MP25, n=7) or 30mmHg (MP30, n=7) using a novel MP system for hypothermic pulsatile perfusion. After MP preservation, the contralateral kidneys were removed and the preserved kidneys heterotopically autotransplanted. Ten minutes after reperfusion, the microcirculation was reassessed. Seven days posttransplant, animals were euthanized and the kidney grafts removed for histological analysis. MP using a mean PP of 25mmHg resulted in higher capillary blood flow after reperfusion. In the MP30 group, 6 out of 7 animals survived, whereas in the MP25 group all animals survived. Overall, improvement in recovery of renal function and a better preservation of structural integrity were seen in the MP25 group compared to the MP30 group. Using a novel system for hypothermic MP, a mean PP of 25mmHg is preferred over a mean PP of 30mmHg.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.