Abstract

Size-exclusion chromatography (SEC) hyphenated to pyrolysis-gas chromatography (Py-GC) has been demonstrated as a powerful tool in polymer analysis. A main limitation to the wider application of the method are the long second-dimension Py-GC analysis times, resulting in limited first-dimension sampling and/or long overall run times. Therefore, we set out to develop an online hyphenated SEC×Py-MS/FID method, removing the GC separation and allowing for a drastically reduced second-dimension analysis time compared to SEC-Py-GC. The pyrolysis method had a cycle time of 1.31 min, which was facilitated by liquid nitrogen cooling of the programmable temperature vaporizer (PTV) used for pyrolysis. The developed method featured no molar mass discrimination for masses above ±1.3 kDa, rendering it applicable to most commercial polymer systems. The method was demonstrated on multiple samples, including a complex industrial sample, yielding chemical composition heterogeneity and in some cases sequence heterogeneity information over the molar mass distribution.