Abstract

A 67-year critically ill patient suffered from a hypertensive crisis (200 mm Hg) because of a norepinephrine overdose. The overdose occurred when the clinician exchanged an almost-empty syringe and the syringe pump repeatedly reported an error. We hypothesized that an object between the plunger and the syringe driver may have caused the exertion of too much force on the syringe. Testing this hypothesis in vitro showed significant peak dosing errors (up to +572%) but moderate overdose (0.07 mL, +225%) if a clamp was used on the intravenous infusion line and a large overdose (0.8 mL, +2700%) if no clamp was used. Clamping and awareness are advised.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.