Abstract

Two cases that developed a delayed hyperopic shift in refraction following implantation of a single-piece hydrophilic intraocular lens (IOL) are described. The haptics of the Akreos Adapt IOL were flexed anteriorly by capsular contraction, leaving a marked gap between the optic and the anterior capsule. A third case that had marked capsule phimosis and similar anterior flexion of the haptics but with a stable refraction is also described. In this case, the anterior and posterior leaves of the capsule fused peripherally and the IOL optic position was normal. The effective power of an IOL depends on the distance between the apex of the cornea and the center of the optic. Capsule contraction without fusion of the peripheral capsule can make the haptics of this IOL design flex anteriorly with posterior movement of the optic and a hyperopic shift in refraction. No author has a financial or proprietary interest in any material or method mentioned.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.