Abstract

Ninety-six patients were implanted with the Omnifit-HA femoral stem and the Omnifit-PSL porous-coated dual-radius acetabular shell (Stryker, Mahwah, NJ) and liners that were sterilized by γ-radiation in air. The mean age at the time of surgery was 52 years (range, 27-78 years), and male/female ratio was 54:42. The mean follow-up was 10.3 years (range, 7-12 years). The survivorship of the stem and cup was 100% and 90%, respectively. Four cups were revised for aseptic loosening, and 6 liners were exchanged. The mean polyethylene wear rate was 0.24 mm/y. Lysis was noted in 23% of the cups. No patient developed femoral lysis distally. This hydroxyapatite-coated stem continues to perform well and appears to protect against the migration of wear debris along the femoral stem.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.