Abstract
Hydrophone has been widely used in testing ultrasonic field distribution. But the hydrophone has a certain directivity, which may bring error in the actual acoustic field measurements. So it is necessary to meet some requirements about the directivity of hydrophone. In the measurement, the spatial directivities of needle hydrophone, membrane hydrophone and robust hydrophone is measured and a comparison about measured data of HIFU acoustic field with them is given. Experimental results indicate that the asymmetric special directivity of hydrophone is due to manufacturing processes and other reasons. As a result, it is not sufficient to test the directivity of hydrophone along one axis. It is necessary to adjust the direction of the hydrophone when characterizing the HIFU field.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.