Abstract

Surfaces of medical grade silicone rubber (Q7-4750, Dow Corning) were modified by repeated (six times) RF plasma treatments using various discharge gases: oxygen, argon, carbon dioxide, and ammonia. The treated samples were stored for a period of 3 months in ambient air, water, or liquid nitrogen. Subsequently, the temporal behavior of the effects of the plasma treatment on the physicochemical surface properties of the silicone rubber was investigated using water contact angle measurements and X-ray photoelectron spectroscopy (XPS). Hydrophobic recovery during 3 months storage in ambient air was considerable and nearly complete for all four plasmas used. Hydrophobic recovery was almost completely suppressed during storage in liquid nitrogen, and only a minor increase of around 10° in advancing water contact angle was observed for all four plasma treatments. Also during storage of treated samples in water, hydrophobic recovery was minimal and initiated again by returning the treated samples to ambient air....

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.