Abstract

The principal accuracy limitation of a hydrogen maser is the uncertainty in the determination of the hydrogen maser wail shift. In this paper, we describe in detail the experiments on hydrogen maser wall shift and introduce briefly works on hydrogen masers in the Shanghai Bureau of Metrology. In experiments, we changed five different size storage bulbs by. turns. These bulbs were all coated in the same way with a locally made F4 coating. Results of the experiments are K(40°C) = -293 ± 17 mHzcm and a(40°C) = (-17 ± 2) × 10-30C-1. The unperturbed hyperfine frequency of the ground state of the hydrogen atomf0-0 was obtained by comparing five hydrogen masers to Loran C navigation system signals for one month. The average value in respect to TAI isf0-0 = 1420 405 751.768 ± 0.002 Hz. This value is closely in agreement with values of f0-0, which have been published by other authors since 1970.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.