Abstract

BackgroundFew studies have assessed hydrogel spacer shrinkage during external-beam radiation therapy following brachytherapy for localized high-risk prostate cancer. This case presentation evaluated the changes in hydrogel spacer appearance by magnetic resonance imaging during external-beam radiation therapy after brachytherapy for prostate cancer and analyzed the effect of this shrinkage on the dose distribution in four cases.Case presentationIn all cases, we implanted 125I sources using a modified peripheral loading pattern for seed placement. The prescribed dose for each implant was 110 Gy. After delivering the sources, a hydrogel spacer was injected. All cases underwent external-beam radiation therapy approximately 1–2 months after brachytherapy. The prescribed dose of external-beam radiation therapy was 45 Gy in 1.8-Gy fractions. Magnetic resonance imaging was performed for evaluation on the day following seed implantation (baseline), at external-beam radiation therapy planning, and during external-beam radiation therapy. The median hydrogel spacer volume was 16.2 (range 10.9–17.7) cc at baseline, 14.4 (range, 9.4–16.1) cc at external-beam radiation therapy planning, and 7.1 (range, 2.0–11.4) cc during external-beam radiation therapy. The hydrogel spacer volume during external-beam radiation therapy was significantly lower than that at external-beam radiation therapy planning. The rectum V60–80 (rectal volume receiving at least 60–80% of the prescribed dose of external-beam radiation therapy) during external-beam radiation therapy was significantly higher than that at external-beam radiation therapy planning.ConclusionsThe potential reduction in hydrogel spacer size during external-beam radiation therapy following brachytherapy can lead to unexpected irradiation to the rectum. This case presentation would be helpful for similar cases.

Highlights

  • Few studies have assessed hydrogel spacer shrinkage during external-beam radiation therapy following brachytherapy for localized high-risk prostate cancer

  • Hydrogel spacers expand the perirectal space as an injected liquid and polymerize into a soft absorbable spacer to increase the distance between the rectum and prostate, reducing the effects of irradiation to the rectum

  • Planning target volume (PTV), planning target volume; BT, brachytherapy; EBRT, external-beam radiation therapy. a Rectum VX defined as the rectum volume receiving at least X% of the prescribed dose of EBRT

Read more

Summary

Conclusions

The potential reduction in hydrogel spacer size during external-beam radiation therapy following brachytherapy can lead to unexpected irradiation to the rectum.

Background
Evaluation during EBRT
Discussion and conclusions