Abstract

The hydrodynamic function and leaflet dynamics of second generation porcine valves prepared with low- or zero-pressure fixation have been studied and compared to first generation porcine bioprostheses, bileaflet, and tilting disc mechanical valves. The Carpentier-Edwards Supra-Annular and Hancock II valves showed lower pressure drops than the Medtronic Intact valve and first generation porcine valves, and comparable overall energy losses to mechanical valves at normal cardiac outputs. Only the zero-pressure fixed Intact valve showed synchronous leaflet opening. Delayed leaflet opening and high opening pressures were found in both low- and high-pressure fixed porcine valves. All porcine bioprostheses showed high open leaflet bending strains. Fixation of valve leaflets with "near zero" pressure fixation and a more physiological neutral geometry is necessary to ensure synchronous leaflet opening at low flows and a reduction in commissural bending strains.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.