Abstract

A new anaesthetic gas scavenging system is described. The resistance of the system appears to be low. When input gas flow rate in the system was zero or 150 litre min-1, misuse of the system produced pressure changes at its collecting points of 26 Pa subatmospheric and 630 Pa above atmospheric pressure, respectively. Suggestions to develop the system and increase its safety are presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.