Abstract

AbstractThis paper reports the results of an optimisation study for a procedure to determine the total selenium and its inorganic species, Se(IV) and Se(VI) using atomic absorption spectrometry combined with hydride generation and in-situ trapping of the analyte on the inner walls of the graphite tube. With the use of the proposed modification, a detection limit (3σ) of 0.018 ng/ml is achieved. This paper presents exemplary results, according to the proposed procedure, for selenium determination in samples of marine water. The concentrations of selenium in the samples ranged from <0.02 ng/ml to 0.16ng/ml of Se(IV) and from <0.02 ng/ml to 0.10 ng/ml of Se(VI).

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call