Abstract
We investigated the effects of the intra-aortic balloon pump (IABP) on endocardial viability ratio (EVR), cardiac output (CO), end-systolic (V(es)) and end-diastolic (V(ed)) ventricular volumes, total coronary blood flow (TCBF), and ventricular energetics (external work [EW], pressure-volume area [PVA]) under different ventricular (E(max) and diastolic stiffness) and circulatory (arterial compliance) parameters. We derived a hybrid model from a computational model, which is based on merging computational and hydraulic submodels. The lumped parameter computational submodel consists of left and right hearts and systemic, pulmonary, and coronary circulations. The hydraulic submodel includes part of the systemic arterial circulation, essentially a silicone rubber tube representing the aorta, which contains a 40-mL IAB. EVR, CO, V(es), and V(ed), TCBF and ventricular energetics (EW, PVA) were analyzed against the ranges of left ventricular E(max) (0.3-0.5-1 mm Hg/cm(3)) and diastolic stiffness V(stiffness) (≈0.08 and ≈0.3 mm Hg/cm(3), obtained by changing diastolic stiffness constant) and systemic arterial compliance (1.8-2.5 cm(3)/mm Hg). All experiments were performed comparing the selected variables before and during IABP assistance. Increasing E(maxl) from 0.5 to 2 mm Hg/cm(3) resulted in IABP assistance producing lower percentage changes in the selected variables. The changes in ventricular diastolic stiffness strongly influence both absolute value of EVR and its variations during IABP (71 and 65% for lower and higher arterial compliance, respectively). V(ed) and V(es) changes are rather small but higher for lower E(max) and higher V(stiffness). Lower E(max) and higher V(stiffness) resulted in higher TCBF and CO during IABP assistance (∼35 and 10%, respectively). The use of this hybrid model allows for testing real devices in realistic, stable, and repeatable circulatory conditions. Specifically, the presented results show that IABP performance is dependent, at least in part, on left ventricular filling, ejection characteristics, and arterial compliance. It is possible in this way to simulate patient-specific conditions and predict the IABP performance at different values of the circulatory or ventricular parameters. Further work is required to study the conditions for heart recovery modeling, baroreceptor controls, and physiological feedbacks.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.