Abstract
In this Letter, a hybrid frequency-time spectrograph combining a tunable optical filter and a dispersive element is presented for measurement of the spectral properties of the two-photon state. In comparison with the previous single-photon spectrograph utilizing the dispersive Fourier transformation (DFT) technique, this method is advanced since it avoids the need for additional wavelength calibration and the electronic laser trigger for coincidence measurement; therefore, its application is extended to continuous wave (CW) pumped two-photon sources. The achievable precision of the spectrum measurement has also been discussed in theory and demonstrated experimentally with a CW pumped periodically poled lithium niobate (PPLN) waveguide-based spontaneous parametric down-conversion photon source. Such a device is expected to be a versatile tool for the characterization of the frequency entangled two-photon state.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.