Abstract

Hybrid digital subtraction angiography was performed after modification of a DSA system. Alternate high voltage (110–120 kV) and low voltage (65–70 kV) were obtained within short time intervals of 60–70 ms with a high voltage switching generator. Experimental phantom studies revealed the exposure dose per single image of hybrid subtraction was approximately 30% of that of conventional DSA. The SNR of hybrid subtraction was improved by increasing the exposure dose and application of postprocessing programs. Clinical application of intraarterial and intravenous DSA revealed that soft tissue artifacts were removed surprisingly well, but image quality decreased because of lowered SNR. Frame integration and matched filtering improved image quality of hybrid subtraction.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.