Abstract
ABSTRACT Hybrid copper–cobalt hexacyanoferrate, a new member of hexacyanoferrate family, was electrodeposited on a glassy carbon electrode and used as an electrochemical sensor for hydrogen peroxide. The modified electrode could electrocatalyze both reduction and oxidation of hydrogen peroxide in neutral solution. When cathodic catalytic current was used, the sensor exhibited a linear response in hydrogen peroxide concentration range of 2.3 × 10−3 ~8.1 × 10−7 mol/L and the detection limit of 6.6 × 10−8 mol/L. The sensor exhibited high electrocatalysis efficiency towards the reduction of hydrogen peroxide and had good stability and excellent anti-interference ability due to the low applied potential of 0.02 V.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.