Abstract

Robotic minimally invasive surgery has been a subject of intense research and development over the last three decades, due to the clinical advantages it holds for patients and doctors alike. Particularly for drug delivery mechanisms, higher precision and the ability to follow complex trajectories in three dimensions (3D), has led to interest in flexible, steerable needles such as the programmable bevel-tip needle (PBN). Steering in 3D, however, holds practical challenges for surgeons, as interfaces are traditionally designed for straight line paths. This work presents a pilot study undertaken to evaluate a novel human-machine visual interface for the steering of a robotic PBN, where both qualitative evaluation of the interface and quantitative evaluation of the performance of the subjects in following a 3D path are measured. A series of needle insertions are performed in phantom tissue (gelatin) by the experiment subjects. User could adequately use the system with little training and low workload, and reach the target point at the end of the path with millimeter range accuracy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.